Chassis Queries and Commands 3
nVision Programming Instructions
Chassis Queries and Commands 3
—
CHASSIS QUERIES AND COMMANDS
Queries
X
Automatic Shuto Timer
AO? ..........Returns the auto o time in seconds. For example, 1200 is 20 minutes.
X
Firmware Version
VER? .........Returns the nVision rmware version in a R08XXXX.XX format.
X
Message Store
MSG? .........Returns the data from the chassis message store.
X
Model
MOD? .........Returns the model name of the chassis. NV is the nVision reference recorder, and NL is the nVision Lab Reference.
X
Modules Available
MODSA? ......The available modules are returned as a sum of the following descriptors:
Lower Module = 1
Upper Module = 2
BARO Module = 4
Therefore, a number of 3 returned means that both the lower and upper modules are populated.
X
Recording
REC:STA! ...(Run Tag) Starts a recording on the nVision, bypassing the screen lock if enabled (without unlocking the nVision)
Examples ...REC:STA!
Starts a recording with a tag of “Run # N”, N being the number of recordings on the nVision.
REC:STA!Location1
Starts a recording with a tag of “Location1.” (Allows up to 22 characters.)
If you want to have spaces in the run tag, use:
REC:STA! HEX [Run Tag
*
]
•
Run Tag
*
is the run tag string converted to a hexadecimal represented string
Example ...REC:STA! HEX 4c6f636174696f6e31
Starts a recording with a tag of “Location1.” (Allows up to 22 characters.)
•
If a blank run tag is desired use: REC:STA! HEX 20
REC:STO! ...Stops the current recording run, bypassing the screen lock if enabled (without unlocking the nVision)