Calibrating Gauges 12
FastCalXP Operation Manual
11 Enter gauge information.
The Gauge number is your number for this particular gauge – its identi er. This number is required and it must be unique among all gauges stored in the
database. Choose the gauge number carefully since the number will be used in the future to identify this speci c gauge from, potentially, a large number of
gauges stored in the database.
The Serial number is that which is provided by the manufacturer. The serial number, along with model and manufacturer is a second method of uniquely
describing the gauge.
The Nickname is not r
equired but may provide another useful description for the gauge.
Calibration due is the date this gauge is next scheduled for calibration. When a calibration has been c
ompleted and a certi cate created, this date is up-
dated automatically using the calibration interval speci ed for the model.