Calibrating Gauges 15
FastCalXP Operation Manual
Autodetection
FastCalXP automatically detects Crystal Engineering’s nVision Reference Recorders, XP2i (XP2) digital pressure gauges and 30 Series calibrators. Once the
gauge and the reference have been mounted on the GaugeCalHP and the serial or USB cable has been connected to the reference and the computer’s ap-
propriate port, the autodetection and calibration can proceed. Calibration of a smart transmitter with a HART modem requires that the HART modem be
connected to the computer and the HART modem leads be connected to the power leads of the transmitter. Refer to the Hardware Setup and Connections
sec
tion
and to the GaugeCalHP Operation Manual for help with con guring y
our calibration work station.
To Automatically Detect a New or Previously Used Reference
1 Turn the reference on and click the Perform New As Received Cal button or the Perform As Left Cal button.
FastCalXP will take a few seconds to detect the reference. When a new reference has been detected it can be added to the database and the current calibra-
tion c
erti cate information must be entered by the operator.
2 Enter the Latest calibration certi cate
number and the date on which that Certi cate expires.
If the auto-detected reference exists in the database, FastCalXP will proceed to the Exercise Window or to the Calibration Window.
If the reference is not automatically detected refer to the Frequently Asked Questions.
Adding References Manually
References can be added manually using the Database Manager. Refer to the Managing References on page 42.