FastCalXP Operation Manual
Appendix
APPENDIX A: FASTCALXP PRESSURE UNITS
FastCalXP uses the following standard units and associated conversion factors from PSI. All conversion factors use a gravity constant of 9.80665 m/sec
2
.
1 PSI = 27.6806 inches of water column (water at 4° C [39.2° F])
27.7070 inches of water column (water at 15.6° C [60° F])
27.7292 inches of water column (water at 20° C [68° F])
2.03602 inches of mercury (mercury at 0° C [32° F])
51.7149 millimeters of mercury (mercury at 0° C [32° F])
703.087 millimeters of water column (water at 4° C [39.2° F])
0.070307 kilograms per square centimeter
68.948 millibar
6.8948 kilopascals
0.068948 bar
0.006895 megapascals
APPENDIX B: EXCEL TEMPLATE TAGS
Following is a list of Excel template tags used by FastCalXP to create certi cates using search-and-replace operations.
Tags are always included within chevrons (e.g., <tag>) and they can occur anywhere within a template worksheet. However, FastCalXP will only search and
replace in a worksheet that includes the <CertNum> tag. Bold tags must be present for FastCalXP to  ll elds de ned by other tags in the same group.
Tags marked “Saved in Cert” are values that will remain constant within a certi cate once the certi cate is created, even if the information in the database
is changed at a later date. For example, the gauge number for a given certi cate will not change even if it is changed in the database after the certi cate is
created. Since the system can recreate the certi cate at a later date, saved  elds provide an assurance that the major elements within a certi cate will not be
modi ed after the certi cate is created.
Some of the tags have a number sign (#) as part of the tag. This is to support multiple values of the same type (i.e., multiple references or data points) to be
included in the certi cate. For example, to have a certi cate that has ten indicated pressure points for run 1, you would need to include <ARRun1Point1Indi-
cated> through <ARRun1Point10Indicated> tags in the template. To have a certi cate that has multiple runs i.e. 5 runs, you would need to include <ARRun-
1Point1Indicated> through <ARRun5Point1I0ndicated>. The registry entry value for the following must be at least the same number of runs in the template:
HKEY_CURRENT_USER\Software\Crystal Engineering\FastCalXP\Settings\ExcelMaxRuns Value Data “5”.