Managing the Database 39
FastCalXP Operation Manual
TEST PROFILES
The Pro les editor provides access to all test pro les stored in the database.
The Pro le Details box is also available for creating a new pro le when a new model is created using the New Gauge Wizard.
Name
When creating a new pro le, it will help to make the Name descriptive and speci c, particularly if you will be using several similar pro les for di erent models.
Units of Measure
Use the pull-down list
to specify the correct Units of measure.
Number of Excercise Cycles
Enter a non-zero Number of exercise cycles to remind the operator to exercise the gauges before beginning any calibration that uses this pro le.
To Create a Uniform Series of Pressure Points
1 Click on Ascending/descending points.
2 Specify the minimum and maximum range of the pro le and the number of asc
ending and descending points.
The Pressure list will be lled automatically.